The long-stroke ZX-TDS04T[] (4-mm measurement range) achieves high-precision measurement with the industry's top class of resolution (0.1 μm) and linearity of 0.3 FS max.
The long-stroke ZX-TDS04T[] (4-mm measurement range) achieves high-precision measurement with the industry's top class of resolution (0.1 μm) and linearity of 0.3 FS max.
ZX-TDS10T-[][] long-range type
Resolution: 0.4 mm, Linearity: ±0.5% FS
Since micropressure contact is achieved using the Actuator's own weight, these Sensor Heads are ideal for taking measurements where non-contact Sensors cannot be used, such as on transparent and glossy products or products that are easily scratched or warped
Switching from dial gauges to automatic inspection is a snap with the ZX-TDS10T-V. Meanwhile, the ZX-TDS10T-VL can control air push models in addition to air lift models so that contact force can now be controlled externally.
The ZX-T uses a differential transformer method, so there is no need to change the absolute origin position. As a result, master adjustment and origin calculation do not have to be performed each time operation is started. There is also no need for complex zero point return operations have been eliminated.
The Amplifier automatically displays the measurement distance when it is connected to the Sensor Head. The cable also be extended up to 10 meters with no effect on characteristics.
After the power is turned ON, this display indicates when the Sensor Head has warmed up to its optimum measurement condition.
The ZX-TDS01T Sensor Head (1-mm measurement range) and the ZX-TDS04T Sensor Head (4-mm measurement range) have long service lives thanks to the unique linear ball-bearing structure shown below. Sliding parts move smoothly and the rubber sleeves keep dust out.
Problems caused by excessive pressing force in inappropriate measurement situations can be detected in advance and a signal can be output to prevent malfunction. For example, preventive measures can be taken with a PLC, such as automatically stopping the measurement.
Select the most appropriate Actuator for your application.
Model | ZX-TDA11 | ZX-TDA41 |
---|---|---|
Measurement period | 1 ms | |
Possible average count settings *1 |
1, 16, 32, 64, 128, 256, 512, or 1,024 | |
Linear output *2 |
Current output: 4 to 20 mA/F.S., Max. load resistance: 300 Ω Voltage output: ±4 V (±5 V, 1 to 5 V *3), Output impedance: 100 Ω |
|
Judgement outputs (3 outputs: HIGH/PASS/LOW) |
NPN open-collector outputs, 30 VDC, 30 mA max. Residual voltage: 1.2 V max. |
PNP open-collector outputs, 30 VDC, 30 mA max. Residual voltage: 2 V max. |
Zero reset input, timing input, reset input, judgement output hold input |
ON: Short-circuited with 0-V terminal or 1.5 V or less OFF: Open (leakage current: 0.1 mA max.) |
ON: Supply voltage short-circuited or supply voltage of 1.5 V or less OFF: Open (leakage current: 0.1 mA max.) |
Function |
Measurement value display, Present value/set value/output value display, Display reverse, ECO mode, Number of display digit changes, Sample hold, Peak hold, Bottom hold, peak-to-peak hold, Self-peak hold, Self-bottom hold, Zero reset, Initial reset, Direct threshold value setting, Position teaching, Hysteresis width setting, Timing inputs, Reset input, Judgement output hold input, (A-B) calculations *4, (A+B) calculations *4, Sensor disconnection detection, Zero reset memory, Function lock, Non-measurement setting, Clamp value setting, Scale inversion, Zero reset indicator, Span adjustment, Warming-up display, Pressing force alarm |
|
Indicators |
Judgement indicators: High (orange), pass (green), low (yellow), 7-segment main digital display (red), 7-segment sub-digital display (yellow), power ON (green), zero reset (green), enable (green) |
|
Power supply voltage | 12 to 24 VDC ±10%, Ripple (p-p): 10% max. | |
Current consumption |
140 mA max. (with Sensor connected), For 24-VDC power supply voltage: 140 mA max. (with Sensor connected) |
|
Ambient temperature | Operating and storage: 0 to 50°C (with no icing or condensation) | |
Temperature characteristic | 0.03% F.S./°C | |
Connection method | Prewired (standard cable length: 2 m) | |
Weight (packed state) | Approx. 350 g | |
Materials | Case: PBT (polybutylene terephthalate), Cover: Polycarbonate |
Model | ZX-TDS01T | ZX-TDS04T | ZX-TDS04T-L | |
---|---|---|---|---|
Measurement range | 1 mm | 4 mm | ||
Maximum actuator travel distance | Approx. 1.5 mm | Approx. 5 mm | ||
Resolution *1 | 0.1 μm | |||
Linearity *2 | 0.3% F.S. | |||
Operating force *3 | Approx. 0.7 N | Approx. 0.25 N | ||
Degree of protection (Sensor Head) | IEC60529, IP67 | IEC60529, IP54 | ||
Mechanical durability | 10,000,000 operations min. | |||
Ambient temperature |
Operating: 0°C to 50°C (with no icing or condensation) Storage: -15°C to 60°C (with no icing or condensation) |
|||
Ambient humidity | Operating and storage: 35% to 85% (with no icing or condensation) | |||
Temperature characteristic *4 |
Sensor Head | 0.03% F.S./°C | ||
Preamplifier | 0.01% F.S./°C | |||
Weight (packed state) | Approx. 100 g | |||
Materials | Sensor Head | Stainless steel | ||
Preamplifier | Polycarbonate | |||
Accessories | Instruction manual, Preamplifier Mounting Brackets (ZX-XBT1) |
Model | ZX-TDS10T | ZX-TDS10T-V | ZX-TDS10T-L | ZX-TDS10T-VL | |
---|---|---|---|---|---|
Vacuum retract (VR) and air push (AP) compatible |
No | VR | No | VR/AP | |
Measurement range | 10 mm | ||||
Maximum actuator travel distance | 10.5 mm | ||||
Resolution *1 *5 | 0.4 μm | ||||
Linearity *2 *5 | ±0.5% FS | ||||
Operating force *3 | Approx. 0.7 N | Approx. 0.6 N | Approx. 0.065 N |
Approx. 0.09 to 1.41 N |
|
Air pressure | Vacuum retracting | --- |
-0.55 to -0.70 (bar) |
--- | -0.22 to -0.5 (bar) |
Air push | --- | 0.125 to 2 (bar) | |||
Degree of protection |
Sensor Head | IP65 | IP50 | ||
Preamplifier | IP40 | ||||
Mechanical durability | 10,000,000 operations min. | ||||
Ambient temperature |
Operating: 0 to 50°C (with no icing or condensation), Storage: -10 to 60°C (with no icing or condensation) |
||||
Ambient humidity | Operating and storage: 35% to 85% (with no icing or condensation) | ||||
Temperature characteristic *4 |
Sensor Head | ±0.01% F.S./°C | |||
Preamplifier | ±0.01% F.S./°C | ||||
Vibration resistance |
0.35-mm single amplitude at 10 to 55 Hz for 50 min each in the X, Y, and Z directions |
||||
Shock resistance | 150 m/S2 3 times each in 6 directions (up/down, left/right, and forward/backward) | ||||
Connection method |
Prewired connector (2 m from the Sensor Head to the Preamplifier, 0.2 m from the Preamplifier to the connector) |
||||
Weight (packed state) | Approx. 100 g | ||||
Materials | Sensor Head | Stainless steel | |||
Rubber sleeve | Viton | None | |||
Preamplifier | Polycarbonate | ||||
Mounting Brackets | Stainless steel | Polyamide | |||
Accessories |
Instruction manual, Preamplifier Mounting Brackets (ZX-XBT1), Right- angle Adapter *6 |
last update: November 06, 2012
* Measurement range: 11.2 to 12.2 (TYP)
* Measurement range: 14.9 to 18.9 (TYP)
last update: November 06, 2012
TITLE | DOCUMENT TYPE | SIZE |
ZX-T Data Sheet | Catalog | 965 KB |
---|---|---|
ZX-T Series Smart Sensors (High-precision Contact Type) Operation Manual |
Manual | 2709 KB |
ZX-CAL2 Calculating Unit for Smart Sensors Instruction Sheet (ENG/JPN) |
Manual | 2753K KB |
ZX-TDA[][] Smart Sensors (Contact Displacement Type) Amplifier Unit Instruction Sheet (ENG/JPN) |
Manual | 3246 KB |
ZX-TDS Smart Sensor (High accuracy Contact-type) sensor unit for model ZX-TDA[][] Instruction Sheet (ENG/JPN) |
Manual | 304 KB |
ZX-TDS10[]-[][] (For Model ZX-TDA[][]) Smart Sensor (High accuracy Contact-type) Sensor Unit Instruction Sheet |